首页 | 本学科首页   官方微博 | 高级检索  
     


Assessment of Atrial Capture in Committed Atrioventricular Sequential (DVI) Pacing Systems
Authors:PAUL A. LEVINE ,SAMUEL J. BRODSKY,JEFFREY P. SELTZER &dagger  
Affiliation:Evans Memorial Department of Clinical Research and Section of Cardiology, University Hospital, Boston, Massachusetts
Abstract:Two patients with committed dual unipolar AV sequential pacing systems are presented in whom alrial capture was unable to be demonstrated on the surface electrocardiogram. The bedside examination is proposed as a means of identifying the presence or absence of atrial capture with subsequent confirmation by M-mode echocardiographs and jugular venous pulse tracings. Mode, rate, and output programmabilitymayaidin the evaluation of thesepacing systems. Intentional reduction of the ventricular output to subthreshold levels in pacemakers capable of this degree of programmabilily will permit the rapid bedside confirmation of atrial capture without the need for expensive and time-consuming noninvasive tests.
Keywords:committed AV sequential pacing    atrial pacing    programmability    AVI pacing