Improvement of blood compatibility on cellulose dialysis membrane. 2. Blood compatibility of phospholipid polymer grafted cellulose membrane.
Authors:K Ishihara  R Takayama  N Nakabayashi  K Fukumoto  J Aoki
Institution:Institute for Medical and Dental Engineering, Tokyo Medical and Dental University, Japan.
Abstract:The blood compatibility of a cellulose haemodialysis membrane whose surface was grafted with a methacrylate having a phospholipid polar group, 2-methacryloyloxyethyl phosphorylcholine, was evaluated with attention to platelet adhesion to the membrane surface and complement activation induced by the membrane. When the original cellulose membrane came in contact with platelet-rich plasma for 30 min, numerous platelets adhered to the surface and aggregated. On the other hand, the membrane grafted with 2-methacryloyloxyethyl phosphorylcholine effectively suppressed platelet adhesion and activation. This effect became more pronounced with increasing surface distribution. Especially, the 2-methacryloyloxyethyl phosphorylcholine grafted membranes, whose distribution exceeded 0.27, completely inhibited platelet adhesion, even when the contact time was 180 min. Moreover, the complement activation was also reduced with increased 2-methacryloyloxyethyl phosphorylcholine distribution on the surface of the membrane.