Pyrford钢丝环扎张力带内固定治疗髌骨骨折的生物力学分析及临床研究
引用本文:吴震东,刘丹,黄宰宇,王以进,王勇,张磊,应志豪,王圣斌. Pyrford钢丝环扎张力带内固定治疗髌骨骨折的生物力学分析及临床研究[J]. 中国矫形外科杂志, 2005, 13(14): 1069-1071
作者姓名:吴震东  刘丹  黄宰宇  王以进  王勇  张磊  应志豪  王圣斌
作者单位:1. 解放军第118医院骨科,温州,325000
2. 上海大学上海生物力学工程研究所,上海,201800
摘    要:[目的] 测试Pvrford钢丝环扎加张力带内固定治疗髌骨骨折的生物力学特性,为临床治疗提供理论依据。[方法] 采集新鲜牛尸体膝关节标本,制成髌骨骨折模型,用Pyrford钢丝环扎加张力带固定,并与克氏针张力带、单纯钢丝环扎进行对照比较。[结果]Pyrford钢丝环扎加张力带内固定其强度和刚度相当于传统的8字形克氏针张力带,而且髌骨的应变、位移很小。临床应用Pyrford钢丝环扎加张力带内固定治疗髌骨骨折105例,随访6~23个月,优良率达98%。[结论]Pyrford钢丝环扎加张力带固定完全符合髌骨的生物力学性能,且能达到解剖复位,操作简便,固定牢固,适应早期功能锻炼的目的。

关 键 词:内固定治疗 髌骨骨折 钢丝环扎张力带 临床研究 钢丝环扎加张力带 分析及 克氏针张力带 张力带固定 生物力学特性 张力带内固定 生物力学性能 早期功能锻炼 膝关节标本 临床治疗 骨折模型 对照比较 临床应用 解剖复位 优良率
修稿时间:2005-04-13

Biomechanical evaluation and clinical application of patella fractures fixed with the combination of cerclage wiring and a tension band ( Pyrford technique )
WU Zhen-dong,LIU Dan,HUANG Zai-yu,et al.. Biomechanical evaluation and clinical application of patella fractures fixed with the combination of cerclage wiring and a tension band ( Pyrford technique )[J]. The Orthopedic Journal of China, 2005, 13(14): 1069-1071
Authors:WU Zhen-dong  LIU Dan  HUANG Zai-yu  et al.
Affiliation:WU Zhen-dong,LIU Dan,HUANG Zai-yu,et al.Department of Orthopaedic surgery,The 118th Hospital of PLA,Wenzhou 325000
Abstract:[Objective]To investigate the probabilities of clinical application of Pyrford technique, and to evaluate biomechanical characteristics of patella fractures fixed with the combination of cerclage wiring and a tension band (Pyrford technique).[Method]By using ox cadaveric knee joints, transverse osteotomies of the patella were performed, and the simulated fractures were fixed with the following techniques: the combination of cerclage wiring and a tension band(Pyrford technique), tension band with kirschner pins, and cerclage with wires alone. The effectiveness of fixation was measured and compared.[Result]Intensity and rigidity of Pyrford technique internal fixation is same to tension band with kirschner pins, and its strain and displacement of patella was very small. Accordingly,105 casas of patellar fracture were treated with figure of Pyrford tension band wiring, the excellent and good results was 98% in follow-up studies ranged from 6 to 23 months.[Conclusion]The patella fractures fixed with the combination of cerclage wiring and a tension band(Pyrford technique) was fit for the biomechanical characteristics of patella, and had advantages such as excellent biomechanical capabilities, anatomical reduction, simple operation, firm fixation,and was suitable to function exercise in early stage.
Keywords:Patella fracture  Internal fixation  Pyrford technique  Biomeehanics