Artificial Pancreas Systems: Closed-Loop Artificial Pancreas Using Subcutaneous Glucose Sensing and Insulin Delivery and a Model Predictive Control Algorithm: The Virginia Experience |
| |
Authors: | William L. Clarke Stacey Anderson Marc Breton Stephen Patek Laurissa Kashmer Boris Kovatchev |
| |
Affiliation: | 1Division of Pediatric Endocrinology, Department of Pediatrics, University of Virginia Health Sciences Center, Charlottesville, Virginia;2Department of Internal Medicine, University of Virginia, Charlottesville, Virginia;3Diabetes Technology Program, University of Virginia, Charlottesville, Virginia;4Department of Systems and Information Engineering, University of Virginia, Charlottesville, Virginia |
| |
Abstract: | BackgroundRecent progress in the development of clinically accurate continuous glucose monitors (CGMs), automated continuous insulin infusion pumps, and control algorithms for calculating insulin doses from CGM data have enabled the development of prototypes of subcutaneous closed-loop systems for controlling blood glucose (BG) levels in type 1 diabetes. The use of a new personalized model predictive control (MPC) algorithm to determine insulin doses to achieve and maintain BG levels between 70 and 140 mg/dl overnight and to control postprandial BG levels is presented.MethodsEight adults with type 1 diabetes were studied twice, once using their personal open-loop systems to control BG overnight and for 4 h following a standardized meal and once using a closed-loop system that utilizes the MPC algorithm to control BG overnight and for 4 h following a standardized meal. Average BG levels, percentage of time within BG target of 70–140 mg/dl, number of hypoglycemia episodes, and postprandial BG excursions during both study periods were compared.ResultsWith closed-loop control, once BG levels achieved the target range (70–140 mg/dl), they remained within that range throughout the night in seven of the eight subjects. One subject developed a BG level of 65 mg/dl, which was signaled by the CGM trend analysis, and the MPC algorithm directed the discontinuance of the insulin infusion. The number of overnight hypoglycemic events was significantly reduced (p = .011) with closed-loop control. Postprandial BG excursions were similar during closed-loop and open-loop controlConclusionModel predictive closed-loop control of BG levels can be achieved overnight and following a standardized breakfast meal. This “artificial pancreas” controls BG levels as effectively as patient-directed open-loop control following a morning meal but is significantly superior to open-loop control in preventing overnight hypoglycemia. |
| |
Keywords: | artificial pancreas closed-loop control continuous glucose monitoring model predictive control algorithms |
|
|