Time for new concepts about measurement of complement activation by cardiopulmonary bypass?
Authors:V Videm  E Fosse  T E Mollnes  P Garred  J L Svennevig
Affiliation:Institute for Experimental Medical Research, Ullevaal Hospital, Oslo, Norway.
Abstract:Fifty-one patients admitted for routine coronary bypass operations were randomized to cardiopulmonary bypass with a membrane oxygenator (Capiox) or a bubbler (Polystan or William Harvey). Complement activation was measured using enzyme immunoassays for concentrations of C3 activation products and the terminal complement complex. From 5.8 to 8.1 arbitrary units (AU)/mL (medians), the plasma concentrations of C3 activation products increased by 119.9 AU/mL (Capiox), 124.6 AU/mL (Polystan), and 79.5 AU/mL (William Harvey) to a peak at closure of the sternum (not significant when related to baseline concentrations). The increase in C3 activation products and baseline C3 activation were linearly correlated (R2 = 0.30; p less than 0.0001). From 5.5 to 6.1 AU/mL, the plasma terminal complement complex concentrations increased by 45.2 AU/mL (Capiox), 15.4 AU/mL (Polystan), and 17.4 AU/mL (William Harvey) to a peak before termination of cardiopulmonary bypass. Maximal terminal (C5-C9) activation was significantly higher in the membrane oxygenator group (p less than 0.0001) and showed no relationship to C3 activation. Measurement of C3 activation only gives no information about C5-C9 activation. At present, terminal complement complex quantitation is probably the best index of C5-C9 activation during cardiopulmonary bypass.