Multicenter evaluation of methods to quantitate human immunodeficiency virus type 1 RNA in seminal plasma
Authors:Fiscus S A,Brambilla D,Coombs R W,Yen-Lieberman B,Bremer J,Kovacs A,Rasheed S,Vahey M,Schutzbank T,Reichelderfer P S  AIDS Clinical Trials Group Genital Secretions Working Group
Affiliation:University of North Carolina, Chapel Hill, North Carolina 27599, USA. fiscussa@med.unc.edu
Abstract:We have evaluated two commercially available kits (AMPLICOR MONITOR [Roche] and NASBA HIV-1 QT or NucliSens HIV-1 QT [Organon Teknika]) and two noncommercial methods for the accurate quantitation of human immunodeficiency virus type 1 (HIV-1) RNA in seminal plasma. The same panels of coded specimens were tested on four separate occasions. Laboratories using the commercial assays employed silica beads to isolate HIV-1 RNA, which removed inhibitory factors sometimes found in seminal plasma. Sensitivities and specificities, respectively, for each assay were as follows: AMPLICOR MONITOR, 100 and 73%; NASBA HIV-1 QT, 84 and 100%; NucliSens HIV-1 QT, 99 and 98%; and noncommercial assays, 91 and 73%. When results from the laboratory that was inexperienced with the silica bead extraction method were excluded from the analysis, specificity for the Roche assay increased to 100%. The commercial assays demonstrated highly reproducible results, with intra-assay standard deviations (measured in log(10) RNA copies/milliliter of seminal plasma) ranging from 0.11 to 0.32; those of the noncommercial assays ranged from 0.12 to 0.75. Differences in mean estimated HIV-1 RNA concentrations were