首页 | 本学科首页   官方微博 | 高级检索  
     

参与IAEA/WHO医用加速器剂量输出比对的结果及分析
引用本文:赵红福,赵志鹏,王志强,程光惠. 参与IAEA/WHO医用加速器剂量输出比对的结果及分析[J]. 中国辐射卫生, 2019, 28(1): 69-71. DOI: 10.13491/j.issn.1004-714x.2019.01.019
作者姓名:赵红福  赵志鹏  王志强  程光惠
作者单位:吉林大学中日联谊医院放射治疗科, 吉林 长春 130033
基金项目:吉林省卫生计生委项目(2014ZC054);吉林省科技厅自然基金(白求恩专项)项目(20160101079JC);吉林大学横向课题(2015373)
摘    要:目的 通过参加国际原子能机构(IAEA)和世界卫生组织(WHO)组织的邮寄热释光剂量计(TLD)国际比对,检查医院加速器剂量输出的准确性。方法 申请IAEA/WHO邮寄TLD剂量核查,获得参与比对资格。每个能量照射两个TLD样品,照射剂量为2 Gy,并填写数据表格。将TLD及数据表邮寄回IAEA实验室。IAEA测量后给出比对结果。结果 6 MV光子束的两个TLD剂量均值为2.00 Gy,平均偏差为0.0%,10 MV光子束的两个TLD剂量均为2.02 Gy,平均偏差为-0.8%。结论 按照IAEA标准的允许偏差±5%考虑,本次比对结果符合比对要求。

关 键 词:热释光  国际比对  绝对剂量  
收稿时间:2018-10-08

Results and analysis of participation in the medical accelerator dose output comparison organized by IAEA/WHO
ZHAO Hongfu,ZHAO Zhipeng,WANG Zhiqiang,CHENG Guanghui. Results and analysis of participation in the medical accelerator dose output comparison organized by IAEA/WHO[J]. Chinese Journal of Radiological Health, 2019, 28(1): 69-71. DOI: 10.13491/j.issn.1004-714x.2019.01.019
Authors:ZHAO Hongfu  ZHAO Zhipeng  WANG Zhiqiang  CHENG Guanghui
Affiliation:Department of Radiation Oncology, China-Japan Union Hospital of Jilin University, Changchun 130033 China
Abstract:Objective To check the accuracy of medical accelerator dose by participating in international comparisons by postal TLD organized by the IAEA/WHO. Methods Apply for the IAEA/WHO postal TLD dose audit to obtain the participation qualification. For each energy condition, irradiate the 2 TLD samples at an irradiation dose of 2 Gy and fill out the data sheet. Mail the TLDs and data sheet back to the IAEA lab, and the results were given after the IAEA measurements. Results The mean measured dose of the 6 MV photon beam was 2.00 Gy with an average deviation of 0.0%, and the doses for 2 TLDs irradiated at 10 MV photon beam were both 2.02 Gy with an average deviation of -0.8%. Conclusion The comparison results met the requirement within allowable deviation of ±5% set by IAEA.
Keywords:TLD  Intercomparison  Absolute Dose  