基于晶体振荡器的MIT系统激励源设计及其实现
引用本文:吕华,秦明新,李世俊,王聪. 基于晶体振荡器的MIT系统激励源设计及其实现[J]. 医学争鸣, 2005, 26(10): 944-947
作者姓名:吕华  秦明新  李世俊  王聪
作者单位:第四军医大学生物医学工程系电子学教研室,陕西,西安,710033;第四军医大学生物医学工程系电子学教研室,陕西,西安,710033;第四军医大学生物医学工程系电子学教研室,陕西,西安,710033;第四军医大学生物医学工程系电子学教研室,陕西,西安,710033
摘    要:目的: 设计一种应用于磁感应断层成像技术(magnetic induction tomography,MIT)系统的10 MHz高性能激励源,以提高MIT系统的测量精度. 方法: 根据高频电子电路设计原理及其电磁屏蔽技术,选用高频率精度和稳定性的恒温晶体振荡器,设计和实现了高性能激励源.结果: 激励源稳定输出10 MHz正弦信号,频率稳定度优于1×10-7水平,杂波抑制和谐波抑制分别达到90 dBc和45 dBc,负载时输出电压峰峰值为12 V,最大输出功率为1.3 W. 结论: 设计的激励源提供了高频率稳定性和高信噪比的激励信号,为高精度的MIT测量奠定了基础.

关 键 词:磁感应断层成像  晶体振荡器  频率稳定性  激励源
文章编号:1000-2790(2005)10-0944-04
修稿时间:2005-01-10

Design and implementation of excitation source based on crystal oscillator for magnetic induction tomography system
Lü Hua,QIN Ming-Xin,LI Shi-jun,WANG Cong. Design and implementation of excitation source based on crystal oscillator for magnetic induction tomography system[J]. Negative, 2005, 26(10): 944-947
Authors:Lü Hua  QIN Ming-Xin  LI Shi-jun  WANG Cong
Abstract:AIM: To improve the measurement precision of magnetic induction tomography (MIT) system by designing a high-performance excitation source.METHODS: On the basis of the principle of the high frequency circuit design and the techniques of electromagnetic shield,an excitation source was designed and implemented by a crystal with high frequency precision and stability.RESULTS: The excitation source output 10 MHz sine wave,with the stability of frequency of 1x10 -7,the cluster suppression of 90 dBc,the harmonics suppression of 45 dBc,the maximum amplitude of 12 V and the maximum output power of 1.3 W.CONCLUSION: Our designed excitation source provides a signal with a higher frequency stability and a higher SNR.The performance of the system lays the basis for high-precision MIT measurement.
Keywords:magnetic induction tomography  crystal oscillator  frequency stability  excitation source