What Range of Programmable AV Delays is Necessary in Antibradycardia DDD Stimulation?
Authors:GEORG H. VON  KNORRE   BRUNO ISMER  WOLFGANG VOSS  MICHAEL PETZSCH  KAMESH PULYA
Affiliation:Department of Cardiology, Clinic for Intemal Medicine, University of Rostock, Germany
Abstract:DDD pacemakers differ considerably in device specific extents of AV delay (AVD) programmability. To demonstrate the requirements of a mean DDD pacemaker patient population optimal AVDs in 200 DDD pacemaker patients (age 8 to 91 years) were estimated by left atrial electrography. The results should help to define an AVD programmability standard. Left atrial electrograms were recorded via a bipolar filtered esophageal lead. The method aims on adjusting the left atrial electrogram to 70 ms prior to the ventricular spike, both during VDD and DDD operation of the pacemaker. In atrial sensed stimulation the optimal AVD varied from 40 to 205 ms (100.5 ± 24.5 ms) and in atrial paced stimulation from 85 to 245 ms (169.1 ±24.5 ms). The difference of the mean values is statistically significant (p < 0.001). The difference between both values in the individual patient, the individual AVD correction time, varied from 0 to 170 ms (68.7 ± 26.6 ms). Thus, from our findings requirements on AV delay programmability standard can be derived: AVDs (1) should have a range from 40 to 250 ms, (2) should be independently programmable during atrial sensed and atrial paced operation, ami (3) should provide as nominal settings 100 ms for atrial sensed and 170 ms for atrial paced stimulation.
Keywords:dual chamber pacemakers    pacemaker AV delay    esophageal electrography