首页 | 本学科首页   官方微博 | 高级检索  
检索        


Doppler ultrasound tracking instrument for monitoring blood flow velocity
Authors:von Krüger Marco Antônio  Evans David H
Institution:Division of Medical Physics, Leicester Warwick Medical School, Leicester, UK.
Abstract:Doppler ultrasound (US) is potentially a valuable method for monitoring changes of blood flow velocity over a period of many minutes or even hours, but is seldom used in this way. One difficulty that may have contributed to this is the problem of maintaining a fixed geometry between the US beam and the blood vessel. A method of improving the success of monitoring might be to actively steer the US beam so as to maintain an adequate signal even when small displacements of the transducer occur. We have designed and built a prototype system for this purpose. The system comprises a continuous-wave phased-array transducer controlled by a purpose-built Doppler unit. The system constantly evaluates the quality of the returning Doppler signal in terms of total power and signal-to-noise ratio (SNR) (evaluated by assessing the quality of derived envelope signals), and steers the ultrasonic beam in a manner so as to improve the signal, should this be necessary. The system was tested in vitro, where the automatic tracking of the Doppler signal doubled the effective beam width of the transducer. Further developments that increase sensitivity and steering range should result in US Doppler systems that are better suited to long-term monitoring.
Keywords:Doppler ultrasound  Blood flow monitoring  Phased-array transducers  Doppler tracking
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号