Clinical study using novel endoscopic system for measuring size of gastrointestinal lesion |
| |
Authors: | Kiyoshi Oka Takeshi Seki Tomohiro Akatsu Takao Wakabayashi Kazuo Inui Junji Yoshino OK Fiber Technology Co.,Ltd. Graduate |
| |
Affiliation: | School of engineering and Resource Science,Akita University;A-tech Co. Ltd.;Department of Internal Medicine,Banbuntane Hotokukai Hospital,Fujita Health University; |
| |
Abstract: | AIM: To verify the performance of a lesion size measurement system through a clinical study.METHODS: Our proposed system, which consists of a conventional endoscope, an optical device, an optical probe, and a personal computer, generates a grid scale to measure the lesion size from an endoscopic image. The width of the grid scale is constantly adjusted according to the distance between the tip of the endoscope and lesion because the lesion size on an endoscopic image changes according to the distance. The shape of the grid scale was corrected to match the distortion of the endoscopic image. The distance was calculated using the amount of laser light reflected from the lesion through an optical probe inserted into the instrument channel of the endoscope. The endoscopist can thus measure the lesion size without contact by comparing the lesion with the size of the grid scale on the endoscopic image. (1) A basic test was performed to verify the relationship between the measurement error eM and the tilt angle of the endoscope; and (2) The sizes of three colon polyps were measured using our system during endoscopy. These sizes were immediately measured by scale after their removal.RESULTS: There was no error at α = 0°. In addition, the values of eM (mean ± SD) were 0.24 ± 0.11 mm (α = 10°), 0.90 ± 0.58 mm (α = 20°) and 2.31 ± 1.41 mm (α = 30°). According to these results, our system has been confirmed to measure accurately when the tilt angle is less than 20°. The measurement error was approximately 1 mm in the clinical study. Therefore, it was concluded that our proposed measurement system was also effective in clinical examinations.CONCLUSION: By combining simple optical equipment with a conventional endoscope, a quick and accurate system for measuring lesion size was established. |
| |
Keywords: | Lesion size Non-contact measurement Endoscope Grid scale Low level laser |
本文献已被 CNKI 等数据库收录! |
|