Use of TEVA resin for the determination of U isotopes in water samples by Q-ICP-MS. |
| |
Authors: | K Tagami S Uchida |
| |
Affiliation: | Environmental and Toxicological Sciences Research Group, National Institute of Radiological Sciences, Anagawa 4-9-1, Inage-ku, Chiba-shi, Chiba 263-8555, Japan. k_tagami@nirs.go.jp |
| |
Abstract: | In order to measure uranium isotopic mass ratio in natural water samples by Q-ICP-MS, an application of TEVA resin (Eichrom) was studied to separate and concentrate U. After being evaporated to dryness, the sample residue was dissolved in 6 M HCl, then, TEVA extraction was carried out. U extracted on the resin could be removed with 20 ml of 1 M HCl (U fraction) when Fe content was lower than 2 mg. U recovery in U fraction showed a negative correlation with Fe content in the samples. |
| |
Keywords: | |
|
|