首页 | 本学科首页   官方微博 | 高级检索  
     

新型数字显示式颈总动脉压迫器
引用本文:孙永达 颜宪秋. 新型数字显示式颈总动脉压迫器[J]. 生物医学工程学杂志, 1997, 14(3): 230-232,236
作者姓名:孙永达 颜宪秋
作者单位:天津大学应用物理系!天津,300072,天津大学应用物理系!天津,300072,天津大学应用物理系!天津,300072,天津大学应用物理系!天津,300072,天津肿瘤医院!天津,300030,海军高等医专!210099,海军高等医专!210099,海军高等医专!210099
基金项目:中国人民解放军海后科研基金
摘    要:本文介绍了新近研制的数显便携式颈总劝脉压迫器的工作原理、电路设计和调试结果,该仪器增强了患侧动脉切术前施行压迫锻炼改善脑侧支循环的效果并为临床实时监测提供客观数据指标。

关 键 词:颈总动脉压迫器 颈总动脉切术 数字显示式 电路

New Digital Carotid Compressor
Sun Yongda, Sun Yongfang, Yuan Lixin, Wang Daoyi, Yan Xianqiu,Zhang Liubao, Chen Jingui, Chen Feng. New Digital Carotid Compressor[J]. Journal of biomedical engineering, 1997, 14(3): 230-232,236
Authors:Sun Yongda   Sun Yongfang   Yuan Lixin   Wang Daoyi   Yan Xianqiu  Zhang Liubao   Chen Jingui   Chen Feng
Affiliation:Tianjin University, Tianjin 300072.
Abstract:Since 1979, many cases of paraganliomas in the neck have been successfully treated in Tianjin Cancer Hospital. All patients who would be subjected to this operation should undergo preoperative compressing exer cises of the carotid artery to promote better circulation of collateral cerebral crossflow. Therefore, a reliable and compact instrument is necessary for the compressing exercises of the carotid artery and the long-term ambulatory monitoring of the compressingeffect.This paper introduces our developed instrument, A djustable and digital Carotid compressor, which is based on the principleof .of..e-o.cillo..t.ic method. During the gradual change in cuff pressure, the amplitude of consecutive arterial volume pulsations associated with pulse pressure shows change characteristically due to the nonlinearity of arterial pressure-volume (P--V) relation. Arterial pressure can be determined by detecting this change in the amplitude, while the arterial volume change can be noninvasively obtained as a function of this transmural pressure, provided that the state of blood flow are determined during this pressure measurement.Considering the simplicity and practicability of this instrument I we have designed a new portable one which consists of pulsepresstlre transducer, signal amplifier, peak voltage locking circuit and liquid crystal displayer. The validity and accuracy of this instrument and a few examples of its clinical application are presented.
Keywords:Digital carotid compressor Carotid resection Arterial pressure