首页 | 本学科首页   官方微博 | 高级检索  
     


An enzyme-immunoassay for myoglobin in human serum and urine. Method development, normal values and application to acute myocardial infarction.
Authors:M J Cloonan  G A Bishop  P D Wilton-Smith  I W Carter  R M Allan  D E Wilcken
Abstract:An enzyme-immunoassay has been developed for the detection of myoglobin in human serum and urine which is specific, accurate, precise, and has a sensitivity of 3 ng/ml. When compared with radioimmunoassay, the enzyme-immunoassay gives markedly similar results. Sera from normal adults had a myoglobin concentration in the range 3-65 ng/ml, and 64% of the same group had detectable myoglobinuria (range 3-11.5 ng/ml). All of 8 patients with definite acute myocardial infarction had raised serum myoglobin levels (range 200-1125 ng/ml) either at admission or 4 h later. Myoglobin concentration returned to normal in 6 patients, and in the remaining 2 patients there was evidence of infarct extension. Urinary myoglobin excretion was variable. One patient with possible acute myocardial infarction had elevated serum myoglobin (413 ng/ml 4 h post admission) and 5 patients with no evidence of infarction had normal levels (15-53 ng/ml). The results suggest that detection of serum myoglobin by enzyme-immunoassay may be a valuable test in the early diagnosis of acute myocardial infarction.
Keywords: