首页 | 本学科首页   官方微博 | 高级检索  
检索        


Safety of Antitachycardia Pacing in Patients with Implantable Cardioverter Defibrillators and Severely Depressed Left Ventricular Function
Authors:ARMIN HEISEL  JÖRG NEUZNER  EWALD HIMMRICH    HEINZ F PITSCHNER  REAS LIEBRICH    JENS JUNG  SEMI SEN
Institution:Innere Medizin III. Universitätskliniken des Saarlandes, Homburg/Saar;Kerckhoffklinik, Bad Nauheim;Medizinische Universitätskliniken. Mainz, Germany
Abstract:The purpose of this study was to investigate the efficacy and safety of antitachycardia pacing (ATP) in third-generation implantable Cardioverter defibrillators (ICDs) for terminating spontaneously occurring ventricular tachycardias (VTs) in patients with severely depressed left ventricular (LV) function. Ninety-one patients with active ATP were followed for 16 ± 13 months. During this period, 775 VT episodes occurring in 36 patients were treated by ATP. The patients were divided into two groups according to their LV ejection fraction (LVEF): group A with LVEF ± 30% (n = 20), and group B with LVEF ± 30% (n = 16). There were no differences between both groups in age, gender, underlying heart disease, indication for ICD therapy, or drug therapy. The VT rates were comparable (group A: 183 ± 16 beats/min; group B: 180 ± 21 beats/min; P = NS). Eighty-three percent of all episodes (n = 332) in group A and 93% of the VTs (n = 443) in group B were ATP terminated (P ± 0.01). Ten percent of VTs in group A were accelerated by ATP into the ventricular fibrillation zone versus 2% in group B (P ± 0.01). The individual termination rate and acceleration rate per patient were comparable in both groups. All VT episodes unresponsive to ATP were converted by backup shocks. The efficacy of first-shock therapy was similar in both groups (group A: 89%; group B: 97%; P = NS). The proportion of patients who needed at least one backup shock for unsuccessful ATP was comparable in both groups (group A: 65%; group B: 56%; P= NS). We conclude that ATP is effective and safe in patients with recurrent VTs and severely depressed LV function, and it can be safely programmed in this group of patients to minimize the use of shock therapy.
Keywords:antitachycardia pacing  implantable cardioverter defibrillator  left ventricular function
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号