Tear Film Break-up Time Evaluated by Real-Time Hartmann-Shack Wavefront Sensing |
| |
Authors: | Toshifumi Mihashi Yoko Hirohara Shizuka Koh Sayuri Ninomiya Naoyuki Maeda Takashi Fujikado |
| |
Affiliation: | (1) Department of Applied Visual Science, Osaka University Graduate School of Medicine, Osaka, Japan;(2) Department of Ophthalmology, Osaka University Graduate School of Medicine, Osaka, Japan;(3) Technical Research Institute, Topcon Corporation, Tokyo, Japan;(4) Department of Applied Visual Science, Osaka University Graduate School of Medicine, Room G4, 2-2 Yamadaoka, Suita, Osaka 565-0871, Japan |
| |
Abstract: | Purpose To develop a novel method for measuring tear break-up time (BUT) by continuously measuring higher order wavefront aberrations using a Hartmann-Shack (H-S) aberrometer. Methods Wavefront aberrations up to the sixth order for a 4-mm pupil were measured in six eyes of six normal subjects using an H-S aberrometer. The aberrometry was performed once every second for up to 50?s under topical anesthesia. The sequence of the points was divided into two stages, and the BUT was defined as the wavefront border of the stages (WFBUT). The WFBUT was compared with the BUT by the conventional fluorescein method (FLBUT) or by the disruption of placid-ring images (PLBUT). Results In three eyes, the WFBUT (average, 10.9 ± 1.1?s) was shorter than the PLBUT (average, 17.5 ± 2.5?s), while in the other three eyes the WFBUT (average, 21.5 ± 2.2?s) was longer than PLBUT (average, 8.1 ± 2.2?s). The FLBUT was shorter than WFBUT or PLBUT in most cases. Conclusions The tear break-up time measured by H-S aberrometer may be used as a practical measure of tear break-up.?Jpn J Ophthalmol 2006;50:85–89 © Japanese Ophthalmological Society 2006 |
| |
Keywords: | aberration break-up time Hartmann-Shack aberrometer tear film |
本文献已被 PubMed SpringerLink 等数据库收录! |
|