首页 | 本学科首页   官方微博 | 高级检索  
检索        


Substitution of carbonate buffer by water for IgG immobilization in enzyme linked immunosorbent assay
Authors:Shrivastav Tulsidas G  Basu Anupam  Kariya Kiran P
Institution:Department of Reproductive Biomedicine, National Institute of Health and Family Welfare, Munirka, New Delhi, India. nihfw@delnet.ren.nic.in
Abstract:The first step of enzyme linked immunosorbent assay (ELISA), namely, adsorption of antigen or antibody to the plastic microtiter well plate, was studied as a function of insolubility of IgG in water. Immobilization efficiency was assessed in terms of number of wells coated per milliliter of primary antiserum. We have compared different coating/immobilization protocols, i.e., direct and indirect immobilization of primary antibody to the plastic microtiter well plate using carbonate buffer and phosphate buffer with glutaraldehyde. We have observed efficient coating when the immobilization of primary antibody through an immunobridge technique was performed, where water was used as a coating medium. It gave a higher number of wells coated per milliliter of anti-serum (primary or secondary) than other compared coating protocols and it allowed the use of serum (non-immune) and anti-serum (primary and secondary antibody) dilutions, avoiding the need for gamma-globulin purification from normal and immunized serum.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏