Extracorporeal ventriculoatrial shunt with the use of one-way ball valve |
| |
Authors: | Kubo Shigeki Takimoto Hiroshi Hosoi Kazuki Toyota Shingo Takakura Shuji Hayashi Yasuhiro Ueno Masato Morisako Toshitaka Karasawa Jun Ninaga Hideo Yoshimine Toshiki |
| |
Affiliation: | Department of Neurosurgery, Osaka Neurological Institute, 2-6-23 Shounai, Takara-machi, Toyonaka-city, Osaka 561-0836, Japan. |
| |
Abstract: | We developed a simple system of an "extracorporeal" ventriculoatrial (VA) shunt using a one-way ball valve (Acty valve II, Kaneka Medix) to release the patient from postoperative constraint during the ventricular drainage. The system is constructed in such a way that the ventricular drainage tube is connected to the central venous catheter via a one-way valve. The CSF is regulated by using the valve and is diverted into the systemic circulation as in the conventional ventriculoatrial shunt. After 2 or 3 weeks of CSF diversion through the extracorporeal VA shunt, a ventriculoperitoneal shunt is placed if hydrocephalus is apparent by temporary occlusion of the system. We applied this system to 4 patients with hydrocephalus, and we found it useful and free from adverse effects. The patient was freed from physical constraint involved in conventional ventricular drainage and an effective program of early rehabilitation was able to be started. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|