首页 | 本学科首页   官方微博 | 高级检索  
检索        


Preliminary Studies with an Implantable Multimodal A-V Pacemaker for Reciprocating Atrioventricular Tachycardias
Authors:AGUSTIN CASTELLANOS  HARVEY L WAXMAN  FEDERICO MOLEIRO  BAROUH V BERKOVITS  RUEY J SUNG
Institution:Division of Cardiology, Department of Medicine, University of Miami School of Medicine, Miami, Florida
Abstract:The efficacy of a recently developed programmable implantable pacemaker which can be used in the management of supraventricular tachycardias (with, or without, associated bradyarrhythmias) was evaluated in the cardiovascular laboratory and at the bedside in the coronary care unit. Although this pacemaker functions primarily in a DVI-DVO mode, it was also made (iatrogenically) to operate in a DOO mode. The study was performed in six patients with reciprocating atrioventricular tachycardias (four of the long P-R--short R-P type and two of the short P-R--long R-P type). Both DVI-DVO and DOO pacing abolished all episodes of tachycardia in five of the patients. However, DOO pacing was also capable of initiating episodes of tachycardia in two of the six patients. The events observed in two of the patients raise the possibility that (reversed) ventriculoatrial sequential pacing may be required in some tachycardias of the short P-R--long R-P type. This preliminary study suggests that programmability of pacing modes, rates, and intervals is a desirable feature that pacemakers used to manage drug-resistant tachycardias should have.
Keywords:multimodal pacemaker  programmable pacemaker  DVJ-DVO pacing  DOO pacing  reciprocating A-V tachycardias