首页 | 本学科首页   官方微博 | 高级检索  
     


MR angiography without subtraction
Authors:Y Zur  X Zou  L J Neuringer
Affiliation:Francis Bitter National Magnet Laboratory, Massachusetts Institute of Technology, Cambridge 02139.
Abstract:A new NMR method for producing angiograms with strong suppression of the static spin signal and without the need for subtraction is described and demonstrated. A velocity-selective pulse sequence was implemented whereby the magnetization of all stationary spins is driven to the -z axis and is not detected, while maximizing the signal intensity of the moving spins. A theory of the method is presented and gives good agreement with experimental results obtained on a flow phantom. It is shown theoretically and experimentally that high-quality velocity-independent angiograms of the head and neck can be obtained with strong suppression of static spin signal when TR approximately T1. The method can be extended to produce three-dimensional angiograms.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号