首页 | 本学科首页   官方微博 | 高级检索  
     


Continued stabilization of trabecular metal tibial monoblock total knee arthroplasty components at 5 years-measured with radiostereometric analysis
Authors:Wilson David A J  Richardson Glen  Hennigar Allan W  Dunbar Michael J
Affiliation:School of Biomedical Engineering, Dalhousie University, Halifax, Canada. wilsondaj@gmail.com
Abstract:

Background and purpose

The trabecular metal tibial monoblock component (TM) is a relatively new option available for total knee arthroplasty. We have previously reported a large degree of early migration of the trabecular metal component in a subset of patients. These implants all appeared to stabilize at 2 years. We now present 5-year RSA results of the TM and compare them with those of the NexGen Option Stemmed cemented tibial component (Zimmer, Warsaw IN).

Patients and methods

70 patients with osteoarthritis were randomized to receive either the TM implant or the cemented component. RSA examination was done postoperatively and at 6 months, 1 year, 2 years, and 5 years. RSA outcomes were translations, rotations, and maximum total point motion (MTPM) of the components. MTPM values were used to classify implants as “at risk” or “stable”.

Results

At the 5-year follow-up, 45 patients were available for analysis. There were 27 in the TM group and 18 in the cemented group. MTPM values were similar in the 2 groups (p = 0.9). The TM components had significantly greater subsidence than the cemented components (p = 0.001). The proportion of “at risk” components at 5 years was 2 of 18 in the cemented group and 0 of 27 in the TM group (p = 0.2).

Interpretation

In the previous 2-year report, we expressed our uncertainty concerning the long-term stability of the TM implant due to the high initial migration seen in some cases. Here, we report stability of this implant up to 5 years in all cases. The implant appears to achieve solid fixation despite high levels of migration initially.Swedish Knee Arthoplasty Register data suggest that uncemented tibial components have worse survivorship than cemented components (Knutson and Robertsson 2010). However, more recent data from Australia have suggested that modern uncemented designs have a similar cumulative revision rate to that of cemented components (Graves et al. 2004). With recent improvements in biomaterials, the potential for improved longevity using uncemented implants is once again being explored. Trabecular metal (Zimmer, Warsaw, IN) is a new material available for use in uncemented total knee arthroplasty. It is a porous biomaterial with morphology and mechanical properties resembling those of trabecular bone (Bobyn et al. 1999a,b, Zardiackas et al. 2001, Levine et al. 2006, Balla et al. 2010). In a previous paper, we presented the 2-year implant migration results of the trabecular metal tibial monoblock component using radiostereometric analysis (RSA) (Dunbar et al. 2009). The results of that study showed high initial migration in 9 out of 28 of the tibial components, with apparent stabilization occurring after 1 year in all cases. These results reflected the results of a similar RSA study on the same implant, performed at another center (Henricson et al. 2008). In addition, in the 2-year report we also documented apparent deformation of the tibial base plate, which occurred in 5 of the cases of high migration (Dunbar et al. 2009). The implications of the high degree of migration and plate deformation for long-term survival are not known. In a subset of the patients enrolled in the study, we also performed RSA for measurement of inducible displacement between 2 and 4 years. The results of this study showed that the trabecular metal component had the lowest inducible displacement ever reported in the literature, indicating excellent stability (Wilson et al. 2010). The results of these 3 studies supported the hypothesis that these components were achieving adequate bone in-growth for long-term survival (Henricson et al. 2008, Dunbar et al. 2009, Wilson et al. 2010). However, longer follow-up is necessary to determine whether the early stability of these implants is durable. In this paper, we present the 5-year longitudinal RSA results from the original cohort of patients randomized to receive either the Nexgen LPS monoblock (trabecular metal) tibial component (Zimmer) or the cemented NexGen Option Stemmed tibial component (Zimmer) (Dunbar et al. 2009).
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号