首页 | 本学科首页   官方微博 | 高级检索  
     


Digital Infrared Fundus Reflectance
Authors:Samuel Packer  Karen Schneider  Hong-Zen Lin  Michael Feldman
Affiliation:1. Ocular Tumor Service, Division of Ophthalmology, North Shore University Hospital, Manhasset, NY;2. Department of Ophthalmology, Cornell University Medical College, New York
Abstract:An infrared sensor was inserted at the film plane of a fundus camera. The signal was visualized on an oscilloscope. In this manner we measured infrared reflectance from the surface of the fundus. The purpose was to characterize choroidal malignant melanomas more reliably than is done with infrared color translation photography. Control lesions were choroidal nevi, metastatic tumors, and disciform macular degenerations. Correlations were made with radioactive phosphorus (32P) uptake, fluorescein angiography, and histopathologic findings. Several cases are presented, one in which this new method of infrared detection was the first diagnostic test to detect the spread of a choroidal melanoma. The simplicity of this technique and its increased accuracy justify the needed further refinements.
Keywords:choroidal melanoma  choroidal nevi  fundus reflectance  infrared  melanoma diagnosis
本文献已被 ScienceDirect 等数据库收录!