首页 | 本学科首页   官方微博 | 高级检索  
     


Implantation of a Subcutaneous Lead Array in Combination with a Transvenous Defibrillation Electrode via a Single Infraclavicular Incision
Authors:JOHN G. KALL  DOUGLAS KOPP  VASSYL LONCHYNA  BRADFORD BLAKEMAN  CHRISTOPHER CADMAN  MARK O'CONNOR  CHARLES KINDER  JAMES GILKERSON    ROBERT AVERY    DAVID WILBER
Affiliation:Electrophysiology Laboratory, Section of Cardiology, Department of Medicine, Loyola University Medical Center, Maywood, Illinois;Department of Cardiothoracic Surgery, Loyola University Medical Center, Maywood, Illinois;Cardiac Pacemakers, Inc., St. Paul, Minnesota
Abstract:Occasional patients have excessive defibrillation energy requirements despite appropriate transvenous defibrillation lead position and modification of defibrillation waveform and configuration. Preliminary data suggest that use of subcutaneous defibrillation electrode arrays with nonthoracotomy systems is associated with a substantial reduction in defibrillation threshold. The current operative approach to subcutaneous lead array implantation involves the use of a separate left chest incision. We present two cases in which implantation of a subcutaneous lead array in combination with a transvenous defibrillation electrode was performed via a single infraclavicular incision and associated with a reduction in defibrillation threshold. Such an approach simplifies implantation and avoids the potential morbidity of the additional incision required of a left lateral chest approach.
Keywords:transvenous defibrillation    nonthoracotomy defibrillation    implantable cardioverterdefibrillator    ventricular fibrillation    subcutaneous defibrillation electrode