A new design for a compact centrifugal blood pump with a magnetically levitated rotor |
| |
Authors: | Asama Junichi Shinshi Tadahiko Hoshi Hideo Takatani Setsuo Shimokohbe Akira |
| |
Institution: | Tokyo Institute of Technology, Interdisciplinary Graduate School of Science and Engineering, Yokohama, Japan. |
| |
Abstract: | A compact centrifugal blood pump has been developed using a radial magnetic bearing with a two-degree of freedom active control. The proposed magnetic bearing exhibits high stiffness, even in passively controlled directions, and low power consumption because a permanent magnet, incorporated with the rotor, suspends its weight. The rotor is driven by a Lorentz force type of built-in motor, avoiding mechanical friction and material wear. The built-in motor is designed to generate only rotational torque, without radial and axial attractive forces on the rotor, leading to low power consumption by the magnetic bearing. The fabricated centrifugal pump measured 65 mm in diameter and 45 mm in height and weighed 0.36 kg. In the closed loop circuit filled with water, the pump provided a flow rate of 4.5 L/min at 2,400 rpm against a pressure head of 100 mm Hg. Total power consumption at that point was 18 W, including 2 W required for magnetic levitation, with a total efficiency of 5.7%. The experimental results showed that the design of the compact magnetic bearing was feasible and effective for use in a centrifugal blood pump. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|