节氧鼻导管的研制及其在氧疗中的节氧效率的观察
引用本文:俞红霞 林江涛. 节氧鼻导管的研制及其在氧疗中的节氧效率的观察[J]. 中华结核和呼吸杂志, 1999, 22(10): 598-601
作者姓名:俞红霞 林江涛
作者单位:中日友好医院呼吸内科
摘    要:目的 研制节氧鼻导管,并探讨其在慢性阻塞性肺疾病(COPD) 患者中的节氧效率。方法 (1)利用吸气相给氧呼气相贮氧的原理,设计并制作节氧鼻导管;(2) 氧疗效果评定:在不同的氧流量(0-5、1-0、1-5、2-0 L/min)下,采用随机、平行、交叉、对照方法,对节氧鼻导管及普通鼻导管氧疗2 小时前后的动脉血气值进行分析,并监测氧疗后不同时间(1、3 、5、10、15、30、60、120 min)的经皮血氧饱和度(SpO2);(3)节氧效率评定:在不同的氧流量(0-5、1-0、1-5、2-0、2-5、3-0 L/min)下,测定采用节氧鼻导管、国外同类产品、普通鼻导管氧疗5 分钟前后的SpO2 值。结果 (1) 在不同的氧流量下,节氧鼻导管组氧疗2 小时后,PaO2 与SaO2 值均较普通鼻导管组高(P<0-05)。(2) 在不同的氧流量下节氧效率分别为:氧流量0-5 L/min 时节氧鼻导管的节氧效率为普通鼻导管的2-6 倍;1-0 L/min 时为1-8 倍;1-5 L/min 时为1-8 倍;2-0 L/min 时为1-5 倍。(3) 节氧鼻导管与国外同类产品比较,节氧效率无显著差异( P> 0-05)。结论 节氧鼻导管的节氧效

关 键 词:节氧鼻导管  肺疾病  阻塞性  氧疗

The study of oxygen-conserving nasal cannula and evaluation of it's oxygen saving ratio in treatment]
YU Hongxia,LIN Jiangtao,ZENG Meiying. The study of oxygen-conserving nasal cannula and evaluation of it's oxygen saving ratio in treatment][J]. Chinese journal of tuberculosis and respiratory diseases, 1999, 22(10): 598-601
Authors:YU Hongxia  LIN Jiangtao  ZENG Meiying
Affiliation:Department of Respiratory Diseases, China-Japan Friendship Hospital, Beijing 100029.
Abstract:OBJECTIVE: To study the oxygen-conserving nasal cannula(CNC), and to evaluate it's oxygen saving benefit ratio in COPD patients. METHODS: (1) CNC was designed as a device which conserved oxygen during exhalation and given oxygen during inhalation; (2) Evaluation of the effect of oxygen therapy: blood analysis was measured at different flows of oxygen (0.5, 1.0, 1.5, 2.0 L/min) before and after oxygen therapy. The test is randomized; Oxygen saturation(SPO2) was measured at 1, 2, 3, 5, 10, 15, 30, 60, 120 minutes by pulse oximeter; (3) Evaluation of the oxygen saving benefit ratio: SPO2 was measured at 0.5, 1.0, 1.5, 2.0, 2.5, 3.0 L/min using CNC, the standard nasal cannula (SCN) and the new CNC. RESULTS: (1) During two hours of oxygen therapy, the values of PaO2 and SaO2 in the CNC group were always higher than those in the SNC group (P < 0.05); (2) When the supply flow to CNC was set at 0.5 L/min, 1.0 L/min, 1.5 L/min, 2.0 L/min seperately, the oxygen saving benefit ratio of CNC to SNC was 2.6:1, 1.8:1, 1.8:1 and 1.5:1 respectively. (3) The oxygen saving benefit ratio was comparable between CNC made in China and that made in foreign country. CONCLUSIONS: It was confirmed that CNC saved oxygen at different flows of oxygen, and the oxygen saving benefit ratio was higher at lower flows than that at higher flows.
Keywords:Oxygen conserving nasal cannula Pulmonary diseases   obstructive Oxygen therapy