自动持续气道正压系统诊断睡眠呼吸暂停综合征的临床评价
引用本文:王浩彦,杨天顺,曹克英. 自动持续气道正压系统诊断睡眠呼吸暂停综合征的临床评价[J]. 中华结核和呼吸杂志, 1999, 22(6): 344-346
作者姓名:王浩彦  杨天顺  曹克英
作者单位:首都医科大学附属北京红十字朝阳医院!100020,Dept.ofMedicine,UniversityofSydney,Australia,Dept.ofMedicine,UniversityofSydney,Australia,Dept.ofMedicine,UniversityofSydney,Australia
摘    要:目的评价简易的自动持续气道正压系统(AutoSet)在临床上诊断睡眠呼吸暂停综合征(SAS)的应用价值。方法120例患者同时进行AutoSet和多导睡眠图(PSG)的监测研究,比较两种方法的呼吸暂停指数(AI)及呼吸暂停+低通气指数(AHI)。结果两种方法的AI及AHI呈显著相关(r=0.88,P<0.001;r=0.93,P<0.001)。AutoSet上显示的呼吸紊乱指数(RII)与PSG的觉醒指数(Ai)呈显著相关(r=0.72,P<0.001)。如以AHI≥5、10、15、20次/小时为诊断标准,AntoSet诊断SAS的敏感性分别为94%、92%、91%、91%,特异性分别为80%、85%、92%、91%。结论AutoSet系统是一个诊断SAS敏感和有价值的工具,尤其对AHI≥30次/小时的SAS患者,其诊断价值更大,但对于轻、中度患者有一定的误差。

关 键 词:睡眠呼吸暂停综合征  多导睡眠仪  自动持续气道正压系统

Comparison of AutoSet with polysomnography in the diagnosis of sleep apnea syndrome
WANG Haoyan, YANG Tianshun, CAO Keying,et al.. Comparison of AutoSet with polysomnography in the diagnosis of sleep apnea syndrome[J]. Chinese journal of tuberculosis and respiratory diseases, 1999, 22(6): 344-346
Authors:WANG Haoyan   YANG Tianshun   CAO Keying  et al.
Affiliation:Beijing Red Cross, Chao Yang Hospital, Capital University of Medical Science, Beijing 100020.
Abstract:OBJECTIVE: To evaluate the validation of the AutoSet system in the diagnosis of sleep apnea syndrome(SAS). METHODS: One hundred and twenty patients were studied simultaneously performed both with the AutoSet and the polysomnography(PSG). The apnea index(AI) and apneas + hypopneas index (AHI) were compared between the two methods by Bland and Altmen method. RESULTS: A significant correlation was found between the AHI assessed by the AutoSet and by the PSG (r = 0.92, P < 0.001). The respiratory irregularity index (RII) of the AutoSet was well correlated with the arousal index (Ai) from the PSG (r = 0.72, P < 0.001). The sensitivity and specificity of the AutoSet in diagnosing sleep apnea were 94% and 80% for AHI > or = 5/hour, 92% and 85% for AHI > or = 10/hour, 91% and 92% for AHI > or = 15/hour, 91% and 91% for AHI > or = 20/hour. CONCLUSIONS: The AutoSet system is a sensitive and valuable tool for identifying patients with sleep apnea and can make a diagnosis of sleep apnea for these patients with AHI > or = 30/hour, but has some limits especially for mild to moderate sleep apnea.
Keywords:Sleep apnea syndrome Polysomnography AutoSet