Improvement of small-field penumbra and dose distributions on a 4 MV accelerator |
| |
Authors: | Ravinder Nath A.G. Agostinelli C.E. Gignac R.J. Schulz |
| |
Affiliation: | Department of Therapeutic Radiology, Yale University School of Medicine, 3333 Cedar Street, New Haven, CT 06510 USA |
| |
Abstract: | The penumbra and dose distributions for small X ray fields from a 4 MV accelerator, designed for 80 cm source-to-surface distance (SSD) but used at 100 cm SSD, were observed to be similar to those obtained from cobalt-60 units. Penumbra was reduced and flatter isodose curves were obtained by the addition of tungsten trimmers at a distance of 48 cm from the X ray source. This stratagem improves the definition of irradiation volume of small fields which are common in the treatment of lesions in the head and neck. |
| |
Keywords: | Radiotherapy accelerators Dose distributions High energy X rays |