首页 | 本学科首页   官方微博 | 高级检索  
检索        


The use of microimplants in orthodontic anchorage.
Authors:Chung-Ho Chen  Chao-San Chang  Chi-Hsin Hsieh  Yu-Chuan Tseng  Yee-Shyong Shen  I-Yueh Huang  Chia-Fu Yang  Chun-Ming Chen
Institution:Department of Oral and Maxillofacial Surgery, Kaohsiung Medical University, Kaohsiung, Taiwan.
Abstract:PURPOSE: Various types of temporary implants have been introduced to serve as orthodontic anchorage. The hypothesis of this study is that microimplants of 1.2 mm diameter can be used as orthodontic anchors, and that their success is related to their length. The aim of this study is to determine the incidence of anchor retention after orthodontic force application for moving teeth, and to determine the relationship of microimplant length to retention rate. METHODS: Fifty-nine microimplants (diameter: 1.2 mm) were placed in 29 patients as orthodontic anchorages. After 2 weeks of microimplant placement, a force of 100 to 200 g was loaded with an elastometric chain or NiTi coil spring. Risk factors were characterized as to why a microimplant may fail, and Fisher's exact test was used for statistical analysis. RESULTS: Nine microimplants were removed and the overall success rate was 84.7%. Exploring the causes for failure, we found significant differences between the length of microimplants and success rate; 6 mm was 72.2% and 8 mm was 90.2%. CONCLUSIONS: The results suggest that microimplants are suited as an alternative orthodontic anchorage. We recommend that 8-mm microimplants are preferable to 6-mm.