Pulsed ultrasound Doppler velocimetry in the assessment of microvascular hemodynamics
Authors:W F Blair  T D Brown  E R Greene
Affiliation:Department of Orthopaedic Surgery, University of Iowa, Iowa City 52242.
Abstract:Pulsed ultrasonic Doppler velocimetry (20 MHz) (PUDVM) has evolved considerably in the last 10 years. Engineering development has resulted in a computer-controlled vessel-scanning instrument whose backscattered frequency shift spectra are analyzed using fast Fourier transforms (FFT). Benchtop and theoretic studies indicate accurate (error less than 5%) velocity and volumetric flow rate measurements in vessels with a lumen diameter as small as 1.2 mm. Clinical application of the PUDVM has provided transcutaneous measurements of blood flow variables in normal human digital arteries. Experimental application to arteries 1.0-1.5 mm has provided information on the hemodynamic effects of topical vasodilators, standard microarteriorrhaphy, variations in microvascular technique, interpositional grafts, and early wound repair. With improving computer capabilities and technical modifications, the PUDVM will be an increasingly important tool in clinical and experimental microsurgery.
Keywords:Hemodynamics  Ultrasound  Microsurgery  Kappa Delta Award