首页 | 本学科首页   官方微博 | 高级检索  
     


Finite-Element Modelling of the Acoustic Input Admittance of the Newborn Ear Canal and Middle Ear
Authors:Hamid Motallebzadeh  Nima Maftoon  Jacob Pitaro  W. Robert J. Funnell  Sam J. Daniel
Affiliation:1.Department of Biomedical Engineering,McGill University,Montréal,Canada;2.Division of Otolaryngology—Head and Neck Surgery,Montréal Children’s Hospital,Montréal,Canada;3.Department of Otolaryngology—Head and Neck Surgery,McGill University,Montréal,Canada;4.Department of Pediatric Surgery,McGill University,Montréal,Canada
Abstract:Admittance measurement is a promising tool for evaluating the status of the middle ear in newborns. However, the newborn ear is anatomically very different from the adult one, and the acoustic input admittance is different than in adults. To aid in understanding the differences, a finite-element model of the newborn ear canal and middle ear was developed and its behaviour was studied for frequencies up to 2000 Hz. Material properties were taken from previous measurements and estimates. The simulation results were within the range of clinical admittance measurements made in newborns. Sensitivity analyses of the material properties show that in the canal model, the maximum admittance and the frequency at which that maximum admittance occurs are affected mainly by the stiffness parameter; in the middle-ear model, the damping is as important as the stiffness in influencing the maximum admittance magnitude but its effect on the corresponding frequency is negligible. Scaling up the geometries increases the admittance magnitude and shifts the resonances to lower frequencies. The results suggest that admittance measurements can provide more information about the condition of the middle ear when made at multiple frequencies around its resonance.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号