Effect of Heart Rate and Coronary Calcification on the Diagnostic Accuracy of the Dual-Source CT Coronary Angiography in Patients with Suspected Coronary Artery Disease |
| |
Authors: | Lingdong Meng Lianqun Cui Yuntao Cheng Xiaoyan Wu Yuansheng Tang Yong Wang Fayun Xu |
| |
Affiliation: | 1Shandong Province Ji''nan 4th People''s Hospital, Ji''nan 250031, China.;2Department of Cardiology, Shandong Provincial Hospital, Ji''nan 250021, China.;3School of Medicine, Shandong University, Ji''nan 250012, China. |
| |
Abstract: | ObjectiveTo evaluate the diagnostic accuracy of a dual-source computed tomography (DSCT) coronary angiography, with a particular focus on the effect of heart rate and calcifications.Materials and MethodsOne hundred and nine patients with suspected coronary disease were divided into 2 groups according to a mean heart rate (< 70 bpm and ≥ 70 bpm) and into 3 groups according to the mean Agatston calcium scores (≤ 100, 101-400, and > 400). Next, the effect of heart rate and calcification on the accuracy of coronary artery stenosis detection was analyzed by using an invasive coronary angiography as a reference standard. Coronary segments of less than 1.5 mm in diameter in an American Heart Association (AHA) 15-segment model were independently assessed.ResultsThe mean heart rate during the scan was 71.8 bpm, whereas the mean Agatston score was 226.5. Of the 1,588 segments examined, 1,533 (97%) were assessable. A total of 17 patients had calcium scores above 400 Agatston U, whereas 50 had heart rates ≥ 70 bpm. Overall the sensitivity, specificity, positive predictive values (PPV) and negative predictive values (NPV) for significant stenoses were: 95%, 91%, 65%, and 99% (by segment), respectively and 97%, 90%, 81%, and 91% (by artery), respectively (n = 475). Heart rate showed no significant impact on lesion detection; however, vessel calcification did show a significant impact on accuracy of assessment for coronary segments. The specificity, PPV and accuracy were 96%, 80%, and 96% (by segment), respectively for an Agatston score less than 100% and 99%, 96% and 98% (by artery). For an Agatston score of greater to or equal to 400 the specificity, PPV and accuracy were reduced to 79%, 55%, and 83% (by segment), respectively and to 79%, 69%, and 85% (by artery), respectively.ConclusionThe DSCT provides a high rate of accuracy for the detection of significant coronary artery disease, even in patients with high heart rates and evidence of coronary calcification. However, patients with severe coronary calcification (> 400 U) remain a challenge to diagnose. |
| |
Keywords: | Dual-source computed tomography Coronary artery disease Coronary angiography |
|
|