A rapid method for electron beam energy check. |
| |
Authors: | V J Pisciotta X Wu J A Fiedler C F Serago M L Calish-Watzich J G Schwade P V Houdek |
| |
Affiliation: | Department of Radiation Oncology, University of Miami School of Medicine, Florida 33101. |
| |
Abstract: | Assessment of electron beam energy and its long term stability is part of standard quality assurance practice in radiation oncology. Conventional depth-ionization or depth-film density measurements are time consuming both in terms of data acquisition and analysis. A procedure is described utilizing ionization measurements at two energy specific depths. It is based on a linear relationship between electron beam energy and its practical range. Energy shifts within the range covered by the two measurement depths are easily resolved. Within a range of +/- 0.50 MeV (+/- 1.30 MeV) around the established mean incident energy of 5.48 MeV (20.39 MeV), the method accuracy is better than 0.10 MeV. |
| |