两种不同检测系统测定血清降钙素原的比较实验研究
引用本文:田 刚,丁俊杰,宋 敏,刘靳波,李光荣.两种不同检测系统测定血清降钙素原的比较实验研究[J].医学信息,2019,0(6):169-171.
作者姓名:田 刚  丁俊杰  宋 敏  刘靳波  李光荣
作者单位:(1.西南医科大学附属医院检验科,四川 泸州 646000;2.江苏三联生物工程有限公司,江苏 无锡 214038)
摘    要:目的 探讨三联生物微阵列化学发光检测系统测定血清降钙素原(PCT)结果的可靠性。方法 以罗氏Cobas e 601电化学发光检测系统为参照系统,三联生物微阵列化学发光检测系统为待评估系统,对血清PCT浓度进行测定并进行统计分析。结果罗氏电化学发光检测系统血清 PCT日内精密度的变异系数(CV)为5.33%~7.88%,日间CV为5.62%~8.56%。三联生物微阵列化学发光检测系统的日内CV为4.93%~7.28%,日间CV为5.47%~8.90%。两种检测系统有较好的相关性(回归方程y=1.174x+0.213,P<0.05),相关系数r2为0.997。在医学决定水平(PCT为0.50 ng/ml和2.0 ng/ml)处的Kappa值分别为0.878(P=4.60×10-7)和0.933(P=7.72×10-8),阳性符合率分别为94.44%和100.00%,阴性符合率分别为93.33%和95.45%。结论 三联生物微阵列化学发光免疫分析法与罗氏电化学发光法测定血清PCT结果相关性和一致性好,可用于临床血清PCT的快速检测。

关 键 词:降钙素原  血清  罗氏电化学发光法  三联生物微阵列化学发光免疫分析法

Comparative Experimental Study on Serum Procalcitonin by Two Different Detection Systems
TIAN Gang,DING Jun-jie,SONG Min,LIU Jin-bo,LI Guang-rong.Comparative Experimental Study on Serum Procalcitonin by Two Different Detection Systems[J].Medical Information,2019,0(6):169-171.
Authors:TIAN Gang  DING Jun-jie  SONG Min  LIU Jin-bo  LI Guang-rong
Institution:(1.Department of Clinical Laboratory,the Affiliated Hospital of Southwest Medical University, Luzhou 646000,Sichuan,China; 2.Jiangsu Sanlian Biological Engineering Co.,Ltd.,Wuxi 214038,Jiangsu,China)
Abstract:Objective To investigate the reliability of the triple biomicroarray chemiluminescence detection system for the determination of serum procalcitonin (PCT) results. Methods The Roche Cobas e 601 electrochemiluminescence detection system was used as the reference system. The triple biomicroarray chemiluminescence detection system was used as the system to be evaluated. The serum PCT concentration was measured and statistically analyzed. Results The coefficient of variation (CV) of intra-day precision of PCT in the Roche electrochemiluminescence detection system was 5.33%~7.88%, and the daytime CV was 5.62%~8.56%. The intra-day CV of the triple biomicroarray chemiluminescence detection system was 4.93%~7.28%, and the daytime CV was 5.47%~8.90%. The two detection systems have a good correlation (regression equation y=1.174x+0.213, P<0.05) and the correlation coefficient r2 is 0.997. The Kappa values at the medical decision level (PCT 0.50 ng/ml and 2.0 ng/ml) were 0.878 (P=4.60×10-7) and 0.933 (P=7.72×10-8), respectively. The positive coincidence rates were 94.44% and 100.00%, the negative coincidence rate was 93.33% and 95.45%.Conclusion The triple biomicroarray chemiluminescence immunoassay and Roche electrochemiluminescence method have good correlation and consistency in the determination of serum PCT results, which can be used for rapid detection of clinical serum PCT.
Keywords:Procalcitonin  Serum  Roche electrochemiluminescence  Triple biomicroarray chemiluminescence immunoassay