首页 | 本学科首页   官方微博 | 高级检索  
     


In vivo percutaneous microwave ablation in kidneys: Correlation with ex vivo data and ablation work
Authors:C. Marcelin  J. Leiner  S. Nasri  F. Petitpierre  Y. Le Bras  M. Yacoub  N. Grenier  J.C. Bernhard  F. Cornelis
Affiliation:1. Service d’imagerie diagnostique et thérapeutique de l’adulte, hôpital Pellegrin, CHU de Bordeaux, place Amélie-Raba-Léon, 33076 Bordeaux, France;2. Service d’anatomopathologie, hôpital Pellegrin, CHU de Bordeaux, place Amélie-Raba-Léon, 33076 Bordeaux, France;3. Service de chirurgie urologique, hôpital Pellegrin, CHU de Bordeaux, place Amélie-Raba-Léon, 33076 Bordeaux, France;4. Service de radiologie, hôpital Tenon, AP–HP, 4, rue de la Chine, 75020 Paris, France
Abstract:

Purpose

To compare diameters of in vivo microwave ablation (MWA) performed in swine kidneys with ex vivo diameters, and to correlate with ablation work (AW), a new metric reflecting total energy delivered.

Material and methods

Eighteen in vivo MWA were performed in 6 swine kidneys successively using one or two antennas (MicroThermX®). Ablation consisted in delivering power (45–120 W) for 5–15 minutes. Ex vivo diameters were provided by the vendors and obtained on bovine liver tissue. AW was defined as the sum of (power)*(time)*(number of antennas) for all phases of an ablation (in kJoules). Kidneys were removed laparoscopically immediately after ablation. After sacrifice, ablations zones were evaluated macroscopically, and maximum diameters of the zones were recorded. Wilcoxon sum rank test and Pearson's correlation were used for comparisons.

Results

For a single antenna (n = 12), the in vivo diameters ranged from 12 to 35 mm, and 15–49 mm for 2 antennas (n = 6). The in vivo diameters remained shorter than ex vivo diameters by 8.6% ± 30.1 on 1 antenna and 11.7% ± 26.5 on 2 antennas (P = 0.31 and 0.44, respectively). AW ranged from 13.5 to 108 kJ. Diameters increased linearly with AW both with 1 and 2 antennas, but only moderate correlations were observed (r = 0.43 [95% confidence interval: ?0.19; 0.81], P = 0.16; and 0.57 [?0.44; 0.95], P = 0.24, respectively).

Conclusion

Although diameters after in vivo renal MWA increased linearly with AW, the moderate correlation and wide standard deviations observed may justify a careful imaging monitoring during treatment delivery and settings adaptation, if needed, for optimal ablation.
Keywords:Microwave ablation  Kidneys  Experimental  Animal model  In vivo
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号