Validation study to determine the accuracy of central blood pressure measurement using the SphygmoCor XCEL cuff device in patients with severe aortic stenosis undergoing transcatheter aortic valve replacement
Authors:Jose M De la Torre Hernndez  Gabriela Veiga Fernandez  Jonathan Brown  Fermin Sainz Laso  Dae&#x;Hyun Lee  Victor Fradejas  Tamara Garcia Camarero  Sammy Elmariah  Ignacio Inglessis  Javier Zueco  Jose A Vazquez de Prada  Eyal Ben&#x;Assa  Elazer R Edelman
Abstract:Central aortic blood pressure could be helpful in the evaluation of patients with aortic stenosis (AS). The SphygmoCor XCEL device estimates central blood pressure (BP) measurement with its easy‐to‐use, operator‐independent procedure. However, this device has not been properly validated against invasive measurement in patients with severe AS. We evaluated the relationship between cuff‐brachial BP, transfer function‐estimated and invasively measured central aortic pressure in patients with severe AS before and after transcatheter aortic valve replacement (TAVR). Agreement between techniques was analyzed and, according to the ARTERY Society recommendations, the minimum acceptable error was a mean difference ± SD ≤5 ± ≤8 mm Hg. A total of 94 patients with AS undergoing TAVR had simultaneous non‐invasive and invasive measurements of central BP before and after the procedure. Before TAVR central systolic BP was in average slightly underestimated, though with wide variability, when using the default calibration of brachial‐cuff SBP (mean difference ± SD, −3 ± 15 mm Hg), and after TAVR the degree of underestimation increased (mean difference ± SD, −9 ± 13 mm Hg). The agreement tended to improve for those patients with low aortic gradient stenosis compared to those with high gradient at baseline (mean difference ± SD, −2 ± 11 mm Hg vs. −4 ± 17, respectively, p = .3). The cuff‐brachial systolic BP yielded numerically lower degree of agreement and weaker correlation with invasive measurements than SphygmoCor XCEL. In patients with severe AS the SphygmoCor XCEL cuff device, despite showing strong correlation, does not meet the ARTERY Society accuracy criteria for non‐invasive measurement of central SBP.
Keywords:aortic stenosis  central aortic blood pressure  hypertension  transcatheter aortic valve replacement
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号