A digital fluoroscopic imaging system for verification during external beam radiotherapy |
| |
Authors: | M Takai |
| |
Affiliation: | Department of Radiology, Hamamatsu University School of Medicine. |
| |
Abstract: | A digital fluoroscopic (DF) imaging system has been constructed to obtain portal images for verification during external beam radiotherapy. The imaging device consists of a fluorescent screen viewed by a highly sensitive video camera through a mirror. The video signal is digitized and processed by an image processor which is linked on-line with a host microcomputer. The image quality of the DF system was compared with that of film for portal images of the Burger phantom and the Alderson anthropomorphic phantom using 10 MV X-rays. Contrast resolution of the DF image integrated for 8.5 sec. was superior to the film resolution, while spatial resolution was slightly inferior. The DF image of the Alderson phantom processed by the adaptive histogram equalization was better in showing anatomical landmarks than the film portal image. The DF image integrated for 1 sec. which is used for movie mode can show patient movement during treatment. |
| |