More than ten years' follow-up of the Hancock porcine bioprosthesis in Japan.
Authors:Y Kawachi  J Tanaka  R Tominaga  K Kinoshita  K Tokunaga
Institution:Division of Cardiovascular Surgery, Faculty of Medicine, Kyushu University, Fukuoka, Japan.
Abstract:From February 1975 through October 1981, 256 Hancock porcine bioprostheses (Johnson & Johnson Cardiovascular, King of Prussia, Pa.) (60 aortic, 169 mitral, and 27 pulmonary/tricuspid position) were implanted in 220 patients (104 male and 116 female, aged 9 to 67 years; mean 43.3) at Kyushu University Hospital in Japan. The procedures include 41 aortic valve replacements, 121 mitral valve replacements, 4 pulmonary valve replacements, 6 tricuspid valve replacements, and 48 combined valve replacements (31 aortic plus mitral, 13 mitral plus tricuspid, and 4 aortic plus mitral plus tricuspid). Hospital mortality was 6.4%. Follow-up was 98% during 8 to 14 (mean 10.5) years. Cumulative follow-up was 1836 patient-years and 2078 valve-years. At 10 years the overall actuarial survival rate, including hospital morality, was 70% +/- 3%, and freedom from valve-related mortality with sudden death was 87% +/- 3%. More than half of the current survivors required no anticoagulant therapy. Freedom from thromboembolism or anticoagulant-related hemorrhage (or both) and prosthetic valve endocarditis was common. Freedom from structural valve failure and reoperation declined more than 9 years after replacement of left-sided heart valves but not after replacement of right-sided heart valves. Sixty-seven patients underwent 68 repeat operations, and there were four deaths (5.9%). The rate of freedom from overall valve-related complications at 10 years was 62% +/- 8% for aortic valve replacement, 53% +/- 5% for mitral valve replacement, 80% +/- 13% for pulmonary/tricuspid valve replacement, and 42% +/- 9% for combined valve replacement. There was a significant difference between pulmonary/tricuspid valve replacement and combined valve replacement (p less than 0.05). The Hancock bioprosthesis is suitable for the replacement of valves in the right side of the heart but not for combined valve replacement.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号