首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical experience with a fiberoptic intracranial pressure monitor
Authors:Jeffrey S. Yablon MD  Howard J. Lantner MD  Thomas M. McCormack MD  Somnath Nair MD  Ellen Barker MSN  Dr Perry Black MD  CM
Affiliation:(1) Department of Neurosurgery, Hahnemann University, Broad and Vine, 19102-1192 Philadelphia, PA
Abstract:Retrospective clinical experience with our first 46 patients monitored with a fiberoptic intracranial pressure device is described. In 43 of 46 patients, the transducer was introduced into brain parenchyma. A ventriculostomy system was used in 3 of 46 patients. The monitoring system was generally characterized by ease of placement and system maintenance and by technical simplicity. Several problems were encountered, including breakage of system components (12%), erroneous readings requiring transducer repositioning (8.6%), epidural hematoma (3.4%), and infection (1.7%). No infections or hematomas occurred in the 3 cases in which the ventriculostomy system was used. Overall, our experience with the Camino intracranial pressure fiberoptic monitoring system confirms previous reports of its favorable features.
Keywords:Complications  Fiberoptic device  Intracranial pressure monitoring
本文献已被 SpringerLink 等数据库收录!