Clinical experience with a fiberoptic intracranial pressure monitor |
| |
Authors: | Jeffrey S. Yablon MD Howard J. Lantner MD Thomas M. McCormack MD Somnath Nair MD Ellen Barker MSN Dr Perry Black MD CM |
| |
Affiliation: | (1) Department of Neurosurgery, Hahnemann University, Broad and Vine, 19102-1192 Philadelphia, PA |
| |
Abstract: | Retrospective clinical experience with our first 46 patients monitored with a fiberoptic intracranial pressure device is described. In 43 of 46 patients, the transducer was introduced into brain parenchyma. A ventriculostomy system was used in 3 of 46 patients. The monitoring system was generally characterized by ease of placement and system maintenance and by technical simplicity. Several problems were encountered, including breakage of system components (12%), erroneous readings requiring transducer repositioning (8.6%), epidural hematoma (3.4%), and infection (1.7%). No infections or hematomas occurred in the 3 cases in which the ventriculostomy system was used. Overall, our experience with the Camino intracranial pressure fiberoptic monitoring system confirms previous reports of its favorable features. |
| |
Keywords: | Complications Fiberoptic device Intracranial pressure monitoring |