Porous cups with and without hydroxylapatite-tricalcium phosphate coating: 23 matched pairs evaluated with radiostereometry
Authors:Thanner J  Kärrholm J  Herberts P  Malchau H
Affiliation:1. Department of Orthopaedic Surgery, Duke University Medical Center, Durham, NC, USA;2. Southern Maryland Orthopaedic and Sports Medicine Center, Leonardtown, MD, USA;1. Department of Surgical and Perioperative Sciences (Orthopedics), Umeå University, Umeå, Sweden;2. Department of Orthopedics, Oslo University Hospital, Ullevål, Oslo, Norway;3. Perth Orthopedic Institute, University of Western Australia, Perth, Australia
Abstract:Migration, wear, and presence of radiolucencies were studied in 23 matched pairs of patients operated with porous-coated acetabular cups with additional screw fixation. All implants had the same type of titanium fiber mesh. In each pair, one of the cups was plasma-sprayed with a coating consisting of 70% hydroxylapatite (HA) and 30% tricalcium phosphate (TCP). Radiostereometric analysis up to 2 years after the operation revealed smaller rotations around the horizontal axis in cups with HA/TCP coating. The migration of the cup center was not significantly influenced. Evaluation of femoral head penetration in 12 of the matched pairs did not reveal any significant difference. Immediately after operation, implants with HA/TCP coating had more central radiolucencies, which, despite minimal migration, disappeared during the follow-up. The clinical results did not differ between the 2 groups. The findings of less tilting and diminishing radiolucencies in the cups with HA/TCP coating suggest a more complete ingrowth of bone and a better sealing of the interface.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号