首页 | 本学科首页   官方微博 | 高级检索  
检索        


The effect of screw insertion angle and thread type on the pullout strength of bone screws in normal and osteoporotic cancellous bone models
Authors:Purvi SD Patel  Duncan ET Shepherd  David WL Hukins
Institution:1. Center for Research on Musculoskeletal Systems (ZESBO), Faculty of Medicine, University of Leipzig, Semmelweisstraße 14, Leipzig 04103, Germany;2. Department of Orthopaedic, Trauma and Plastic Surgery, University of Leipzig, Liebigstraße 20, Haus 4, Leipzig 04103, Germany;3. Faculty of Engineering, University of Applied Sciences Leipzig, Postfach 301166, Leipzig 04251, Germany;1. Department of Biomedical Engineering, National Yang-Ming University, Taipei, Taiwan;2. Department of Orthopaedics, Taipei City Hospital, Taipei, Taiwan;3. Orthopaedic Device Research Center, National Yang-Ming University, Taipei, Taiwan;4. Department of Neurosurgery, Taipei City Hospital, Taipei, Taiwan;5. School of Medicine, National Yang-Ming University, Taipei, Taiwan
Abstract:Screw fixation can be extremely difficult to achieve in osteoporotic (OP) bone because of its low strength. This study determined how pullout strength is affected by placing different bone screws at varying angles in normal and OP bone models. Pullout tests of screws placed axially, and at angles to the pullout axis (ranging from 10° to 40°), were performed in 0.09 g cm?3, 0.16 g cm?3 and 0.32 g cm?3 polyurethane (PU) foam. Two different titanium alloy bone screws were used to test for any effect of thread type (i.e. cancellous or cortical) on the screw pullout strength. The cancellous screw required a significantly higher pullout force than the cortical screw (p < 0.05). For both screws, pullout strength significantly increased with increasing PU foam density (p < 0.05). For screws placed axially, and sometimes at 10°, the observed mechanism of failure was stripping of the internal screw threads generated within the PU foam by screw insertion. For screws inserted at 10°, 20°, 30° and 40°, the resistance to pullout force was observed to be by compression of the PU foam material above the angled screw; clinically, this suggests that compressed OP bone is stronger than unloaded OP bone.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号