首页 | 本学科首页   官方微博 | 高级检索  
     


Eureka-project: ‘totally implantable artificial larynx’: progress report
Authors:Verkerke  Schutte  Mahieu  Van Den Hoogen  De Vries  Geertsema
Abstract:Objective. The European Eureka project ‘Artificial larynx’ aims at realizing an implantable artificial larynx, consisting of artificial vocal folds, artificial epiglottis, tissue connector for tracheal fixation and valve mechanism to switch between breathing and speaking. The last part has been realized. Because its functioning is identical to a tracheostomal valve, it was tested as such. Materials and methods. The device consists of two valves: a speech valve, that will be open during normal breathing and closed by strong expiration and a cough valve that opens during coughing and closes automatically during inhalation. Magnets keep the valves in the breathing position. The magnetic force can be varied. Several prototypes have been tested in vitro and in situ. Results. The prototypes showed good in vitro behaviour. The range of adjustability of speech and cough was wide enough, flow resistance acceptable. This was confirmed by preliminary in vivo experiments. Patients were enthusiastic; hands-free speaking and coughing is possible. Conclusion. Numerous problems have to be solved to produce an artificial larynx. Each solution generated up till now, however, can already be applied to improve the existing rehabilitation process of laryngectomized patients.