Designing long-T2 suppression pulses for ultrashort echo time imaging.
Authors:Peder E Z Larson  Paul T Gurney  Krishna Nayak  Garry E Gold  John M Pauly  Dwight G Nishimura
Affiliation:Magnetic Resonance Systems Research Laboratory, Department of Electrical Engineering, Stanford University, Stanford, California 94305-9510, USA. peder@mrsrl.stanford.edu
Abstract:Ultrashort echo time (UTE) imaging has shown promise as a technique for imaging tissues with T2 values of a few milliseconds or less. These tissues, such as tendons, menisci, and cortical bone, are normally invisible in conventional magnetic resonance imaging techniques but have signal in UTE imaging. They are difficult to visualize because they are often obscured by tissues with longer T2 values. In this article, new long-T2 suppression RF pulses that improve the contrast of short-T2 species are introduced. These pulses are improvements over previous long-T2 suppression pulses that suffered from poor off-resonance characteristics or T1 sensitivity. Short-T2 tissue contrast can also be improved by suppressing fat in some applications. Dual-band long-T2 suppression pulses that additionally suppress fat are also introduced. Simulations, along with phantom and in vivo experiments using 2D and 3D UTE imaging, demonstrate the feasibility, improved contrast, and improved sensitivity of these new long-T2 suppression pulses. The resulting images show predominantly short-T2 species, while most long-T2 species are suppressed.
Keywords:ultrashort echo time (UTE) imaging  long‐T2 suppression  T2 contrast  short‐T2 imaging  RF pulse design
设为首页 | 免责声明 | 关于勤云 | 加入收藏