首页 | 本学科首页   官方微博 | 高级检索  
检索        


Modelling of post-irradiation events in polymer gel dosimeters.
Authors:M Lepage  A K Whittaker  L Rintoul  S A B?ck  C Baldock
Institution:Centre for Medical, Health and Environmental Physics, Queensland University of Technology, Brisbane, Australia.
Abstract:The nuclear magnetic resonance (NMR) spin-spin relaxation time (T2) is related to the radiation-dependent concentration of polymer formed in polymer gel dosimeters manufactured from monomers in an aqueous gelatin matrix. Changes in T2 with time post-irradiation have been reported in the literature but their nature is not fully understood. We investigated those changes with time after irradiation using FT-Raman spectroscopy and the precise determination of T2 at high magnetic field in a polymer gel dosimeter. A model of fast exchange of magnetization taking into account ongoing gelation and strengthening of the gelatin matrix as well as the polymerization of the monomers with time is presented. Published data on the changes of T2 in gelatin gels as a function of post-manufacture time are used and fitted closely by the model presented. The same set of parameters characterizing the variations of T2 in gelatin gels and the increasing concentration of polymer determined from FT-Raman spectroscopy are used successfully in the modelling of irradiated polymer gel dosimeters. Minimal variations in T2 in an irradiated PAG dosimeter are observed after 13 h.
Keywords: