Affiliation: | Department of Bioanalytical Sciences, The Wellcome Research Laboratories, Langley Court, Beckenham, Kent BR3 3BS, UK |
Abstract: | Bioanalytical methods, based on a variety of physico-chemical and biological techniques such as chromatography, immunoassay and mass spectrometry, must be validated prior to and during use to engender confidence in the results generated. The fundamental criteria for assessing the reliability and overall performance of a bioanalytical method are: the evaluation of drug and analyte stability, selectivity, limits of quantification and detection, accuracy, precision, linearity and recovery. The extent to which a method is validated is dependent on its prospective use, the number of samples to be assayed and the use to which the data are put. Specific analytical techniques may require additional validation such as antibody-binding characteristics, peak purity determination, evaluation of matrix effects or structural confirmation of the analyte. Ideally each assay should be cross-validated with a method utilizing a highly specific detector such as a mass spectrometer. Once in use, the performance of the method should be monitored using quality control standards. If a method is set up in another laboratory, the performance of the assay should be monitored with quality control standards sent from the originating laboratory. |