Comparison of scanning equalization and conventional chest radiography
Authors:Wandtke, JC   Plewes, DB
Affiliation:Department of Diagnostic Radiology, University of Rochester Medical Center, NY 14642.
Abstract:A clinical comparison study of scanning equalization radiography (SER) and conventional chest radiography was performed with the latest prototype SER system. Conventional chest radiography was performed at 120 kVp with Lanex regular screens (Eastman Kodak, Rochester, NY) and Kodak Ortho-G or Ortho-C film (Eastman Kodak). The 253 volunteer patients were examined with both techniques. The chest radiographs were interpreted by four radiologists. The study group was composed of 58 normal and 195 abnormal posteroanterior and lateral chest radiographs. In 31 cases there were two major radiologic diagnoses. The number of correct interpretations increased when the SER images were examined, compared with the conventional Ortho-G (chi 2 = 4.17, P less than .05) and conventional Ortho-C (chi 2 = 16.9, P less than .001) radiographs. The overall accuracy of disease detection improved for all radiologists with the SER system. There was no disease category in which the accuracy of interpretation decreased when the SER system was used. The SER system is a clinically reliable method of improving image quality and increasing diagnostic accuracy.