Reproducibility of loading measurements with skin-mounted accelerometers during walking
Authors:Liikavainio Tuomas  Bragge Timo  Hakkarainen Marko  Jurvelin Jukka S  Karjalainen Pasi A  Arokoski Jari P
Affiliation:Department of Physical and Rehabilitation Medicine, Kuopio University Hospital, Kuopio, Finland. liikavai@hytti.uku.fi
Abstract:OBJECTIVE: To examine reproducibility of load measurements with skin-mounted accelerometers (SMAs) during walking. DESIGN: Reliability study. SETTING: A motion analysis laboratory. PARTICIPANTS: Ten healthy young men. INTERVENTIONS: Not applicable. MAIN OUTCOME MEASURES: Two triaxial accelerometers were fixed to the subjects' skin above and below the knee joint. The subjects walked barefoot at their preferred speed and at a constant speed (1.3m/s, +/-5%) in a gait laboratory and along a corridor. The same protocol was repeated over 2 days. Initial peak acceleration (IPA), peak-to-peak (PP) acceleration, and maximal and average acceleration transient rates (ATRs) were calculated. The coefficient of variation (CV) and Pearson linear correlation coefficient were calculated to measure reproducibility of SMA load measurements. RESULTS: IPA and PP acceleration had good interday repeatability (CV <15%). The repeatability of average ATR and maximal ATR parameters was generally not acceptable. The loading variables obtained from ground reaction forces and SMA measurements during gait revealed high linear correlations, indicating that with SMA measurements it is possible to predict certain ground reaction force loading parameters. CONCLUSIONS: SMAs are practical for use in clinical environments to collect acceleration data that may be used to estimate joint loads.
Keywords:Acceleration   Biomechanics   Gait   Reproducibility of results   Rehabilitation
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号