Functional and Biocompatibility Performances of an Integrated Maglev Pump-Oxygenator
Authors:Tao Zhang  Guangming Cheng  †Andrew Koert  Juntao Zhang  †Barry Gellman  G Kwame Yankey  Aditee Satpute  †Kurt A Dasse  ‡Richard J Gilbert  Bartley P Griffith  Zhongjun J Wu
Institution:Artificial Organs Laboratory, Department of Surgery, University of Maryland School of Medicine, Baltimore, MD;;Levitronix LLC, Waltham;and;Department of Mechanical Engineering, Massachusetts Institute of Technology, Cambridge, MA, USA
Abstract:To provide respiratory support for patients with lung failure, a novel compact integrated pump-oxygenator is being developed. The functional and biocompatibility performances of this device are presented. The pump-oxygenator is designed by combining a magnetically levitated pump/rotor with a uniquely configured hollow fiber membrane bundle to create an assembly free, ultracompact, all-in-one system. The hemodynamics, gas transfer and biocompatibility performances of this novel device were investigated both in vitro in a circulatory flow loop and in vivo in an ovine animal model. The in vitro results showed that the device was able to pump blood flow from 2 to 8 L/min against a wide range of pressures and to deliver an oxygen transfer rate more than 300 mL/min at a blood flow of 6 L/min. Blood damage tests demonstrated low hemolysis (normalized index of hemolysis NIH]∼0.04) at a flow rate of 5 L/min against a 100-mm Hg afterload. The data from five animal experiments (4 h to 7 days) demonstrated that the device could bring the venous blood to near fully oxygen-saturated condition (98.6% ± 1.3%). The highest oxygen transfer rate reached 386 mL/min. The gas transfer performance was stable over the study duration for three 7-day animals. There was no indication of blood damage. The plasma free hemoglobin and platelet count were within the normal ranges. No gross thrombus is found on the explanted pump components and fiber surfaces. Both in vitro and in vivo results demonstrated that the newly developed pump-oxygenator can achieve sufficient blood flow and oxygen transfer with excellent biocompatibility.
Keywords:Biocompatibility  Gas transfer  Hemodynamics  Pump-oxygenator  Respiratory support
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号