心声频谱的诊断学价值1、冠心病与正常心声频谱的线性预测分析
引用本文:聂冠群,齐俊瑞,聂金雷,郭秀芹,陈丽,林荣辉,胡旭东,江岩,宣清华,尹光旭,史桂昌,张双荣,陶中达,李键,孙靖. 心声频谱的诊断学价值1、冠心病与正常心声频谱的线性预测分析[J]. 中国生物医学工程学报, 1995, 0(1)
作者姓名:聂冠群  齐俊瑞  聂金雷  郭秀芹  陈丽  林荣辉  胡旭东  江岩  宣清华  尹光旭  史桂昌  张双荣  陶中达  李键  孙靖
作者单位:河北保定人体声学研究所,北京心肺血管医疗研究中心安贞医院,河北省医学科学院,北京海淀医院,中国科学院声学研究所,船泊工业总公司6971厂
摘    要:本文用频率响应不同的三种传声器自69名冠心病人、34名隐匿型冠心病人、40名正常人引导心声信号,用线性预测法脱机分析326个心动周期,共提取出158850个表征心声图象的原始数据,从中经t检验筛选出了在正常与隐匿型冠心病之间,及隐匿型冠心病与冠心病之间有显著差异的特征量。发现用频响低限为0.01Hz的传声器自两组患者胸骨左缘第4肋间导出的心声,其第一心音之末至第二心音之始,以及第三心音之始至下一周期第一心音之始的两个心声段中,0-10Hz频率区间的能量比例显著减少、25-500Hz频率区间的能量比例显著增多、心声频谱的最大谱峰位置与波谱形心位置基本位于可听阈以下的次声带范围内,但均显著向高频方向移动。这些规律可望用于冠心病早期诊断。

关 键 词:冠心病,线性预测分析,心声频谱

DIAGNOSTIC VALUE OF HEART SOUND SPECTRUM:1. LINEAR PREDICTION ANALYSIS OF CORONARY HEARTDISEASE AND NORMAL HEART SOUND
Nie Guan-qun,Qi Jun-rui,Nie Jin-lei,Guo Xiu-qin,Chen Li,Lin Rong-hui. DIAGNOSTIC VALUE OF HEART SOUND SPECTRUM:1. LINEAR PREDICTION ANALYSIS OF CORONARY HEARTDISEASE AND NORMAL HEART SOUND[J]. Chinese Journal of Biomedical Engineering, 1995, 0(1)
Authors:Nie Guan-qun  Qi Jun-rui  Nie Jin-lei  Guo Xiu-qin  Chen Li  Lin Rong-hui
Abstract:Three microphones with different frequency response were used in thisexperiment to lead heart sound signals from 69 cases of coronary heartdisease, 34 cases of latent coronary heart disease , and 40 normal indi-viduals. 158850 original data expressing the phonocardiogram were extra-cted off-line from 326 cardiac cycles with linear prediction analysismethod. Characteristic parameters of significant difference between gro-ups of normal and latent coronary disease, and those between groups oflatent coronary disease and coronary disease were selected by t-testfrom the original data. It was found that when the heart sounds wereled from the left sternal border in the fourth intercostal space of thetwo groups of patienst by the microphone with a lower limit offrequencyresponsc heart at 0.01 Hz, the following variations happened in the twosound segments (from the end of the first heart sound to the beginningof the second heart sound, and from the beginning of the cycle) i thethird heart sound to the beginning of the first heart sound of the nextenergy ratio of frequency between 0-10 Hz was decreased significantly,the energy ratio of frequency between 25-500Hz was increased significa-ntly, both the maximal spectral peak position and spectral centroid pos-ition moving toward the higher frequency direction, though they remai-ned basically in the subsonic area below the audiblc threshold. Thesepattcrns might be nsed in early diagnosis of coronary disease.
Keywords:Coronary disease  Linear prediction analysis  Heart sound spectrum