ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality
Authors:Hong C  Becker C R  Huber A  Schoepf U J  Ohnesorge B  Knez A  Brüning R  Reiser M F
Institution:Institute of Clinical Radiology, Klinikum Grosshadern, University of Munich, Marchioninistrasse 15, 81377 Munich, Germany.
Abstract:PURPOSE: To evaluate the effectiveness of electrocardiographically (ECG)-gated retrospective image reconstruction for multi-detector row computed tomographic (CT) coronary angiography in reducing cardiac motion artifacts and to evaluate the influence of heart rate on cardiac image quality. MATERIALS AND METHODS: Sixty-five patients with different heart rates underwent coronary CT angiography. Raw helical CT data and ECG tracings were combined to retrospectively reconstruct at the defined consecutive z position with a temporal resolution of 250 msec per section. The starting points of the reconstruction were chosen between 30% and 80% of the R-R intervals. The relationships between heart rate, trigger delay, and image quality were analyzed. RESULTS: Optimal image quality was achieved with a 50% trigger delay for the right coronary artery and 60% for the left circumflex coronary artery. Optimal image quality for the left anterior descending coronary artery was equally obtained at 50% and 60% triggering. A significant negative correlation was observed between heart rate and image quality (P <.05). The best image quality was achieved when the heart rate was less than 74.5 beats per minute. CONCLUSION: To achieve high image quality, the heart rate should be sufficiently slow. Selection of appropriate trigger delays and a decreasing heart rate are effective to reduce cardiac motion artifacts.
Keywords:
本文献已被 PubMed 等数据库收录!
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号