Authors:C. S. Arteaga de Castro  V. O. Boer  M. P. Luttje  T. A. van der Velden  A. Bhogal  M. van Vulpen  P. R. Luijten  U. A. van der Heide  D. W. J. Klomp
Affiliation:1. University Medical Center Utrecht, , Utrecht, the Netherlands;2. Netherlands Cancer Center, , Amsterdam, the Netherlands
Abstract:Spectral degradations as a result of temporal field variations are observed in MRSI of the human prostate. Moving organs generate substantial temporal and spatial field fluctuations as a result of susceptibility mismatch with the surrounding tissue (i.e. periodic breathing, cardiac motion or random bowel motion). Nine patients with prostate cancer were scanned with an endorectal coil (ERC) on a 7‐T MR scanner. Temporal B0 field variations were observed with fast dynamic B0 mapping in these patients. Simulations of dynamic B0 corrections were performed using zero‐ to second‐order shim terms. In addition, the temporal B0 variations were applied to simulated MR spectra causing, on average, 15% underestimation of the choline/citrate ratio. Linewidth distortions and frequency shifts (up to 30 and 8 Hz, respectively) were observed. To demonstrate the concept of observing local field fluctuations in real time during MRSI data acquisition, a field probe (FP) tuned and matched for the 19 F frequency was incorporated into the housing of the ERC. The data acquired with the FP were compared with the B0 field map data and used to correct the MRSI datasets retrospectively. The dynamic B0 mapping data showed variations of up to 30 Hz (0.1 ppm) over 72 s at 7 T. The simulated zero‐order corrections, calculated as the root mean square, reduced the standard deviation (SD) of the dynamic variations by an average of 41%. When using second‐order corrections, the reduction in the SD was, on average, 56%. The FP data showed the same variation range as the dynamic B0 data and the variation patterns corresponded. After retrospective correction, the MRSI data showed artifact reduction and improved spectral resolution. B0 variations can degrade the MRSI substantially. The simple incorporation of an FP into an ERC can improve prostate cancer MRSI without prior knowledge of the origin of the dynamic field distortions. Copyright © 2014 John Wiley & Sons, Ltd.
Keywords:field probe  artifact correction  MRSI  7   T  prostate  endorectal coil (ERC)
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号